MAINTENANCE
5-36 31 Dec 13
DE-7200VFDCentrifuge
FAULT CROSS REFERENCE
No. Fault No. Fault No. Fault
2 Auxiliary Input 38 Phase U to Grnd 81-85 Port 1-5 DPI Loss
3 Power Loss 39 Phase V to Grnd 87 IXo VoltageRange
4 UnderVoltage 40 Phase W to Grnd 88 Software Fault
7 Motor Overload 42 Phase VW Short 90 Encoder Quad Err
8 Heatsink OvrTemp 43 Phase UW Short 91 Encoder Loss
9 Trnsistr OvrTemp 48 Params Defaulted 92 Pulse In Loss
12 HW OverCurrent 49 Drive Powerup 93 Hardware Fault
13 Ground Fault 51 Flt QueueCleared 100 Parameter Chksum
15 Load Loss 52 Faults Cleared 101-103 UserSet Chksum
16 Motor Thermistor 55 Cntl Bd Overtemp 104 Pwr Brd Chksum1
17 Input Phase Loss 63 Shear Pin 105 Pwr Brd Chksum2
20 TorqProv Spd Band 64 Drive OverLoad 106 Incompat MCB-PB
21 Output PhaseLoss 69 DP Resistance 107 Replaced MCB-PB
24 Decel Inhibit 70 Power Unit 108 Anlg Cal Chksum
25 OverSpeed Limit 71-75 Port 1-5 Adapter 120 I/O Mismatch
28 See Manual 77 IR Volts Range 121 I/O Comm Loss
29 Analog In Loss 78 FluxAmpsRef Rang 122 I/O Failure
33 Auto Rstrt Tries 79 Excessive Load 130 Hardware Fault
36 SW OverCurrent 80 AutoTune Aborted 131 Hardware Fault
CONTROL COMPONENT INDICATORS
Various indicators are provided to display the operational status of components in the control
cabinet. Figures 5-14 through 5-17 locate and define the indications shown on these components.
Where applicable, corrective actions are included to assist the operator in fault analysis.
Dangerously high voltage is present in the control cabinet. Since opening the control cabinet door
and bypassing the cabinet purge system is required to check the indicators, only trained, qualified
personnel should be permitted to perform these procedures. Use extreme caution to ensure that
the surrounding atmosphere is free of hazardous fumes before opening the cabinet door and
bypassing the purge system.
! ONLY TRAINED, QUALIFIED PERSONNEL SHOULD BE
PERMITTED TO OPEN CONTROL CABINET DOOR WHILE POWER IS
APPLIED.
HAZARD! BE CERTAIN THAT SURROUNDING ATMOSPHERE IS
CLEAR OF ALL POTENTIALLY EXPLOS
IVE GASES BEFORE OPENING